We are unable to set the DC offset in kodiaq standalone mode. There is an option in the config file:
8098 DAC configuration for each channel
(1000 = negative unipolar + offset, 8000=bipolar, F000=positive
register 8098 8000
The value set here does not influence the baseline at all (tried different values). We try different options for:
baseline_mode {int}: 0 - no baselines (baseline from file), 1 - at start of each run
baseline_mode 1
including option 2 in the newer version of kodiaq, which does nothing with the baselines. Options 0 and 2 give a bipolar range (0V = 8000 ADC counts), option 1 sets the DAC offset to unipolar negative (0V = 16200 ADC counts).
Maybe I don't get the config, but does 'determine baseline' here just mean 'measure baseline' or also 'set baseline'? Is it supposed to have such behaviour?
We are unable to set the DC offset in kodiaq standalone mode. There is an option in the config file:
8098 DAC configuration for each channel
(1000 = negative unipolar + offset, 8000=bipolar, F000=positive
register 8098 8000
The value set here does not influence the baseline at all (tried different values). We try different options for:
baseline_mode {int}: 0 - no baselines (baseline from file), 1 - at start of each run
baseline_mode 1
including option 2 in the newer version of kodiaq, which does nothing with the baselines. Options 0 and 2 give a bipolar range (0V = 8000 ADC counts), option 1 sets the DAC offset to unipolar negative (0V = 16200 ADC counts).
Maybe I don't get the config, but does 'determine baseline' here just mean 'measure baseline' or also 'set baseline'? Is it supposed to have such behaviour?